Micro-Metering
Valves
For fine control of fluid flow rates, Micro-Metering Valves can reduce outgoing flow to as low as 3.5 µL/min*. These needle valves are perfect for use with peristaltic pump fluidtransfer applications, mass spectrometry, and fraction collection. Our Micro-Metering Valves can also be used to regulate gas flow in helium sparging lines and as a flow-dependent variable back pressure regulator. For flow independent regulation of back pressure, please see page 149. Flow path materials are PEEK polymer and PTFE. All versions of this valve have 0.020” (0.50 mm) thru-holes.
- Flow rates as low as 3.5 µL/min*
- 1/4-28 flat-bottom and 10-32 coned designs available
- Materials of construction: PEEK, PTFE
* At 1.0 mL/min incoming flow rate with
room temperature water
Back Pressure Considerations
The Micro-Splitter Valves are designed to
work when both effluent flow path pressures
are nearly identical. However, the split flow
path will often have higher back pressure
than the waste flow path, making it hard to
achieve any split flow at all. There are two
possible solutions. Place a back pressure
regulator (page 145) on the waste flow
path that is equal to or slightly greater than
the pressure on the split flow path.
Or, switch the two effluent pathways such
that the split flow pathway is attached to
the “waste” port on the valve and the
waste flow pathway is attached to the
“split” port on the valve. (Please Note: This
second method may result in a loss
of adjustment sensitivity.)
Multi-Column and Detector Systems
Does your work require analyses with
multiple columns and detectors that use
the same mobile phase? If so, install one
of our High Pressure Micro-Splitter Valves
after your injector. A single injection can
then be split to two separate columns and
detector systems, at two different flow rates.
This economical set-up eliminates
the need for an additional pump and
injector valve, while allowing data to be
obtained simultaneously.